Recommended separation distances between portable and mobile RF communication equipment and the
Force EZ-C Series Electrosurgical Generator
The Force EZ-C Series Electrosurgical Generator is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The Customer or the user of the Force EZ-C Series Electrosurgical
Generator can help prevent electromagnetic interferences by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Force EZ-C Series Electrosurgical Generator as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80MHz
d=0.5
√
P
80MHz to 800MHz
d=0.5
√
P
800MHz to 2.5GHz
d=
√
P
0.01
0.05 m
0.05 m
0.1 m
0.1
0.16 m
0.16 m
0.32 m
1
0.5 m
0.5 m
1 m
10
1.6 m
1.6 m
3.2 m
100
5 m
5 m
10 m
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At a 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.