Performance Characteristics
4-6
Force EZ-C Series Service Manual
Input Power
Power Cord Specification
This unit was equipped from the factory with either a 110VAC hospital grade NEMA
5-15 power cord or a 220VAC CEE7/7 power cord. Should the AC power cord need
to be replaced to match another plug configuration, the replacement plug/cable/
receptacle configuration must meet or exceed the following specifications:
100-120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft (5 m)
Plug - minimum 10 A - 125 VAC
Unit receptacle - IEC female, minimum 10 A - 125 VAC
220-240 VAC
Cable - H05VVF3G1.0 VDE, maximum length 15’ (5 meters)
Plug - minimum 6 A - 250VAC
Unit receptacle - IEC female, minimum 6 A - 250VAC
100–120 Volt
220–240 Volt
Maximum VA input:
Idle: 50 VA
Bipolar: 500 VA
Cut: 850 VA
Coag: 500 VA
Maximum VA input:
Idle: 50 VA
Bipolar: 500 VA
Cut: 850 VA
Coag: 500 VA
Full regulation range: 90–135 Vac
Full regulation range: 186–264 Vac
Operating range: 85–140 Vac
Operating range: 170–280 Vac
Mains current (maximum):
Idle: 0.4 A
Bipolar: 4.2 A
Cut: 8.0 A
Coag: 4.2 A
Mains current (maximum):
Idle: 0.2 A
Bipolar: 2.1 A
Cut: 4.0 A
Coag: 2.1 A
Mains line frequency range (nominal):
50 Hz to 60 Hz
Mains line frequency range (nominal):
50 Hz to 60 Hz
Mains fuse: F8.0 A
Mains fuse: T4.0 A
Power plug: 3-prong hospital grade
connector
Power plug: 3-prong locally approved
connector