Responding to System Alarms
T
roubleshooting
Force EZ-C Series Service Manual
7-17
161
163
Dosage error while generator was
activated, where Vsns and Isns stay the
same or decrease while the ECON
increases consistently for about 150 ms.
Dosage error while activating the low 1
coag setting or either of the two high coag
settings.
Do not attempt to use the generator. Record the
number and call the Valleylab Service Center.
164
165
166
Dosage error detected for actual current
values while activating the bipolar mode, a
cut mode, or the low 2 or low 3 coag
setting.
Dosage error detected for actual power
values while activating the bipolar mode, a
cut mode, or the low 2 or low 3 coag
setting.
Dosage error detected for actual voltage
values while activating a cut mode, or the
low 2 or low 3 coag setting.
If the error occurs during calibration, repeat the failing
calibration step. If the number reappears, record the
number and call the Valleylab Service Center.
or
If the error occurs during normal use, do not attempt to
use the generator. Record the number and call the
Valleylab Service Center.
170
Watchdog malfunction. Correct value not
sent from feedback microcontroller.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
171
172
Watchdog malfunction. Correct value not
sent from master microcontroller.
Watchdog malfunction. Master
microcontroller unable to access watchdog
semaphore.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
173
Watchdog malfunction. Feedback
microcontroller unable to access watchdog
semaphore.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
174
Software malfunction.
which_errors
value
outside state range.
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call the
Valleylab Service Center.
180
Internal diagnostics. Master microcontroller
ST Microelectronics RAM check failed.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
181
Internal diagnostics. Feedback
microcontroller ST Microelectronics RAM
check failed.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
182
183
Internal diagnostics. Master microcontroller
dual-port RAM check failed.
Internal diagnostics. Master microcontroller
Page 0 (zero) Flash Memory CRC test
failed.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
Number
Description
Recommended Action