Responding to System Alarms
7-18
Force EZ-C Series Service Manual
184
Internal diagnostics. Master microcontroller
Page F Flash Memory CRC test failed.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
185
Internal diagnostics. Feedback
microcontroller Flash Memory CRC test
failed.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
189
Software malfunction.
display_state
value
outside state range.
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call the
Valleylab Service Center.
191
192
193
194
195
196
197
198
Internal diagnostics. Cut mode buttons
(Pure and/or Blend) may be stuck.
Internal diagnostics. Coag mode buttons
(High and/or Low) may be stuck.
Internal diagnostics. Front panel
Footswitch Selector button may be stuck.
Internal diagnostics. Handswitch or cut
pedal may be stuck.
Internal diagnostics. Handswitch or coag
pedal may be stuck.
Internal diagnostics. Handswitch or cut key
may be stuck.
Internal diagnostics. Handswitch or coag
key may be stuck.
Internal diagnostics. Handswitch or bipolar
footswitch pedal may be stuck.
1.
Turn off, then turn on the generator Do not press
buttons or accessory activation devices during the
self-test.
2.
If the alarm number reappears, disconnect all
accessories. Turn off, then turn on the generator
again.
If the alarm number reappears, record the number
and call the Valleylab Service Center.
199
Internal diagnostics. Master and feedback
microcontrollers are not compatible.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
200
201
Internal diagnostics. Master microcontroller
SEML line may be stuck.
Internal diagnostics. Feedback
microcontroller SEMR line may be stuck.
1.
Replace the Control board. Refer to
Control Board
in Section 8 for instructions.
2.
Calibrate the generator. Refer to Section 6 for
instructions.
If the alarm number reappears, record the number
and call the Valleylab Service Center.
202
Internal diagnostics. Master microcontroller
watchdog test failed.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
203
Internal diagnostics. Feedback
microcontroller watchdog test failed.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
Number
Description
Recommended Action