Low Voltage Power Supply
8-24
Force EZ-C Series Service Manual
Low Voltage Power Supply
Equipment
•
Phillips screwdriver
Remove the Low Voltage Power Supply
1.
Turn off the generator and disconnect the power cord from the wall
receptacle.
2.
Loosen the four screws that secure the cover to the chassis. Lift the cover off
the chassis.
3.
Remove the Control board.
a.
Remove the screw that secures the retention bracket to the left rear heat
sink. Set the screw and bracket aside for reinstallation.
b.
Unlock the connector on the Control board and disconnect the Display
board ribbon cable from the Control board.
c.
Carefully slide the Control board straight up to disconnect it from the
Power Supply/RF board. Set the board aside in an antistatic container for
reinstallation.
4.
Remove the two screws that secure the low voltage power supply to the rear
panel. Save them for reinstallation.
5.
Snap the low voltage power supply off the two standoffs and lift it off the rear
panel.
6.
Disconnect the Power Supply/RF board cables (spade lug and two other
connectors) from the low voltage power supply.
Warning
Electric Shock Hazard
.– To allow stored energy to dissipate after disconnecting
power, wait at least five minutes before replacing parts.
Caution
The generator contains electrostatic-sensitive components. When repairing the
generator, work at a static-control workstation. Wear a grounding strap when
handling electrostatic-sensitive components,
except
when working on an
energized generator. Handle circuit boards by their nonconductive edges. Use an
antistatic container for transport of electrostatic-sensitive components and circuit
boards.