Front Panel
Controls, Indicators,
and
R
ecept
acles
Force EZ-C Series Service Manual
3-5
R.
Footswitch Receptacle, Button, and Indicators
Connect a two-pedal Valleylab monopolar footswitch to this receptacle.
Press the Footswitch Selector button to select bipolar or accessory output.
Use only a Valleylab monopolar footswitch with the Force EZ-C Series
Electrosurgical Generator. Use on an incompatible footswitch may cause
unexpected output.
When the left arrow indicator illuminates green, the footswitch activates the
instrument connected to the Bipolar Instrument receptacle.
When the right arrow indicator illuminates green, the footswitch activates the
instrument connected to the Accessory Instrument receptacle.
The footswitch will not activate an instrument connected to the Monopolar
Instrument receptacle.
S.
Bipolar Instrument Receptacle
You can connect either a handswitching instrument (three-pin connector) or a
footswitching instrument (two-pin connector) to this receptacle.
To activate a bipolar footswitching instrument, you can connect a bipolar
footswitch to the rear panel or a monopolar footswitch to the front panel.
T.
Power Switch
This switch supplies power to the generator.
To turn on the generator, press ( | ).
To turn off the generator, press (
O
).