Performance Characteristics
T
e
chnical S
pecifications
Force EZ-C Series Service Manual
4-7
Low Frequency (50-60 Hz) Leakage Current
High Frequency (RF) Leakage Current
Enclosure source
current, ground open
< 100 µA
Source current, patient
leads, all outputs
Normal polarity, intact ground: < 10 µA
Normal polarity, ground open: < 50 µA
Reverse polarity, ground open: < 50 µA
Sink current at high
line, all inputs
< 20 µA
Bipolar RF leakage
current
< 60 mA
rms
Monopolar RF leakage
current
< 150 mA
rms