Responding to System Alarms
7-20
Force EZ-C Series Service Manual
224
Internal diagnostics. Calibration data
checksum error on feedback
microcontroller.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
225
Internal diagnostics. Feedback
microcontroller CPU test failed.
1.
Replace the Control board. Refer to
Control Board
in Section 8.
2.
Calibrate the generator. Refer to Section 6 for
instructions.
If the alarm number reappears, record the number
and call the Valleylab Service Center.
226
Feedback microcontroller unable to access
FEEDBACK_SEM semaphore.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
230
231
Software malfunction.
keyed_state
value
outside state range.
Software malfunction.
Key_req_mode
value outside state range.
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call the
Valleylab Service Center.
232
Feedback microcontroller unable to access
KEY_ACTIVE_SEM semaphore.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
233
240
241
242
244
245
Software malfunction.
Low_mode
or
High_mode
value in
ac_calc_desired_vals
( ) outside state
range.
Software malfunction.
activation_seq
value outside state range.
Software malfunction.
Feedback
value in
do_cut_feedback
outside state range.
Software malfunction.
feedback
value in
do_cal_feedback
( ) outside state range.
Software malfunction.
disable_seq
value
outside state range.
Software malfunction.
spark_state
value
outside state range.
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call the
Valleylab Service Center.
246
247
Feedback microcontroller unable to access
FEEDBACK_SEM semaphore.
Feedback microcontroller unable to access
ECON_SEM semaphore.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
Number
Description
Recommended Action