Front Panel
Controls, Indicators,
and
R
ecept
acles
Force EZ-C Series Service Manual
3-3
I.
Coag Indicator
When you activate the generator in coag mode, this bar illuminates blue and
an activation tone sounds.
J.
Coag Display
Shows the power setting, in watts, for coag output.
K.
Coag Power Control Knob
To increase (+) the power, turn the knob clockwise.
To decrease (–) the power, turn the knob counterclockwise.
L.
Low (Desiccate) Button
Select to desiccate the area of tissue that is in direct contact with the active
electrode.
M.
High (Fulgurate) Button
Select to fulgurate an area of tissue with a spray of sparks.
N.
REM Alarm Indicator
This indicator illuminates red until you properly apply a REM patient return
electrode to the patient and connect it to the generator. Then, the indicator
illuminates green. (When you connect an electrode without the REM safety
feature, the indicator light does not illuminate.)
If the REM system senses an alarm condition, the indicator flashes red until
you correct the alarm condition – then the indicator illuminates green. (If you
are using a return electrode without the REM safety feature, the red indicator
is extinguished when you correct the alarm condition.)