(6) Volume correction
Change zero shift, span correction and so on.
Fig. 2.2.4-11 Volume correction menu
1. Zero shift
Change zero shift value. Unit is defined by flow/mass/energy unit.
Table 2.2.4-15 Range of zero shift
Default
min.
max.
0.000
-99999.0
999999.0
2. Span correction
Change span correction value.
Table 2.2.4-16 Range of span correction
Default
min.
max.
1.000
0.000
20.000
The correction value is calculated below equation,
q
=
a
x
v
+
b
where,
“
q
” is correction value, “
a
” is span correction, “
v
” is without correction value and “
b
” is
zero shift.
V o l u m e C o r r e c t i o n
1 :
Z e r o S h i f t
0 . 0 0 0 m
3
/ s
2 : S p a n C o r r e c t i o n
1 . 0 0 0
3 : Z e r o C u t
0 . 0 0 0 m
3
/ s
4 : V o l u m e C o r r e c t i o n
R e y n o l d s
# 0 1 : 1 0 0 / S t e e l / 1 M
B a c k
S e l e c t