1-84
2-a. Transducer distance setting
Set distance between transducers on mounting fixture in accordance with the main unit
calculation.
Fig. 1.2.9-9 Message of mounting information Fig. 1.2.9-10 Set transducer distance on fixture
2-b. Set mounting fixture onto the pipe
Wrap the mounting chain around the pipe and hook an end link with the hook knob
arrangement.
Tighten the chain at the other end of the fixture.
1. Roll chain around the pipe. 2.Hook the chain at appropriate length.
3. Tighten chain by knob. 4.Take same procedure on the other side,
then complete.
Fig. 1.2.9-11 Set mounting fixture
M e s s a g e
M o u n t i n g I n f o r m a t i o n
CH1
Distance XX.X mm
Sound-path V-path method
Transducer UP10AST
1
Knob
3
Hook
Chain
2
4
Transducer distance
Slide
Slide