1-11
1.2.2 Connection ports & display indication of the main unit
(1) Setup location
Consider the following conditions when selecting a setup location for the main unit.
a. Place the main unit in a location with an ambient temperature range of -10 to +50 degree C.
Do not place the unit near a heating element and avoid exposure to direct sunlight.
b. Do not place the unit in an area with a corrosive atmosphere.
c. Do not place the unit in a location where interference through electrical devices and power
lines could occur.
d. Please be noted that IP65 can not kept following conditions
•
with AC adaptor connection
•
with USB memory connection
•
without connector cover on the connector
Fig. 1.2.2-1 Sensor connection port
Fig.1.2.2-2 USB & AC adaptor connection port
USB port
AC adaptor
connection port
SENSOR2 (CH2)
connection port
SENSOR1 (CH1)
connection port
Temperature junction box
connection port
Analog output cable
connection port