1-85
2-c. Add couplant and set transducers to mounting fixture
Add silicone grease as acoustic couplant onto surface of transducers.
Then set them into mounting fixture.
1. Add couplant onto surface 2. Set transducer to fixture. 3. Fix transducers by screw.
of transducer as photo.
Fig. 1.2.9-12 Set the transducers into mounting fixture
2-d. Set cables with the transducers and the main unit
Connect cables with the transducers and the main unit.
1. Connect with the transducers. 2. Connect with the main unit.
Fig. 1.2.9-13 Connect with the transducer cables
2
3
Upstream
(Green)
Downstream
(Red)
2
Insert connector
to be straight.
1
Flow
1