1-106
4B-f. Set mounting fixture by magnet type fixture with distance “F-Dist”.
Remove the paper around the pipe at the intended flow measure point.
Position the Transducer-mounting holders by magnet fixture on the pipe and align in
accordance with the lines marked F-Dist, which is the distance between the opposing inside
edges of the transducers.
Magnet will be switched on by turning rotary switch.
Fig. 1.2.9-48 Magnet switch
Fig. 1.2.9-49 Position of the mounting fixtures with the magnets
ON
OFF
ON
OFF
Magnet attachment
Magnet attachment
F-Dist line
Extend Marking Line
Red Line on fixture