1-46
(2) Installation wizard for massmeter with 1path/1pipe or 2path/1pipe
This portable unit can be selectable from “flow meter”, “mass meter” and “heat meter”.
Also “1path / 1pipe (1 location)”, “2path / 1pipe (1 location)” and “1path / 2 pipe (2 locations)” can
be selected on each meter mode.
Here you can see easy-setup procedure (Installation Wizard) for meter type with “1path / 1pipe (1
location)”, “2path / 1pipe (1 location)” for “mass meter”.
If you would like to change each value independently through the advanced setup menu
(3:<Advanced Setting>), please refer to the Chapter 2.
Required parameters
Before started, please confirm following value for input.
•
Diameter or
the circumference of the pipe
•
Thickness of the pipe
•
Material of the pipe
•
Thickness of the lining (if any)
•
Material of the lining (if any)
•
Fluid type
Table 1.2.7-1 Sample of item parameters
Meas. Setting
1path/1pipe or 1path/2pipe
Filename
100/Steel/1M
Pipe Diameter
114.30 mm
Pipe Thickness 4.50 mm
Pipe Material
Carbon Steel
Lining
None
Fluid Type
Water
Sensor Type
UP10AST
Mass Unit
kg/h
Mass Total Unit kg
Note1:
Check “Meter Type” set as “mass meter” on the following directory.
Menu – 7: System – 2: Measuring Setting – 2: Meter Type ; mass meter
Note2:
Check “Path” set as “1path/1pipe” or “2path/1pipe” on the following
directory.
Menu – 7: System – 2: Measuring Setting – 2: Path ; x path / 1 pipe
Summary of Contents for UFP-20
Page 1: ...Doc No KF08 002L Portable Ultrasonic Flowmeter UFP 20 Installation Operation Manual ...
Page 3: ... 2 The attachment of warning label is as follows Fig 1 2 Rear view of main unit Warning label ...
Page 8: ...1 Installation ...
Page 9: ......
Page 17: ......
Page 18: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 130: ...2 Operation ...
Page 131: ......
Page 182: ...3 Other ...
Page 183: ......
Page 193: ...3 8 3 2 4 Dimensions 1 Main Unit Fig 3 2 4 1 Main Unit Hand Strap ...