2-19
(3) Sensor condition
Change the transducer, sound-path and so on.
Fig. 2.2.4-5 Sensor condition menu
1. Type
Press
Select
key (F3 button) and select item using direction button.
Press
Enter
key (F3 button) to confirm the selected item.
Table 2.2.4-7 Transducer type
Transducer type
Recommended diameter
Fluid temperature
UP50AST
13A ~ 50A
-20 ~ 120 deg. C
UP10AST
65A ~ 500A (20A ~ 50A)
-20 ~ 120 deg. C
UP04AST
300A ~ 5000A
-20 ~ 80 deg. C
User Defined
none
none
2. Sound-path
Press
Select
key (F3 button) and select item using direction button.
Press
Enter
key (F3 button) to confirm the selected item.
Table 2.2.4-8 Selectable items of Sound-path
Sound-path
Z-path method (Fig. 2.2.4-6)
V-path method (Fig. 2.2.4-7)
W-path method (Fig. 2.2.4-8)
S e n s o r C o n d i t i o n
1 :
T y p e
U P 1 0 A S T
2 : S o u n d - p a t h
V - p a t h m e t h o d
3 : C a b l e L e n g t h
7 m
# 0 1 : 1 0 0 / S t e e l / 1 M
B a c k
S e l e c t
Note1:
Please select Z-path method with UP10AST, when it is applied for smaller pipe diameter
as less than DN50mm.
Note2:
When it is applied exceeding DN3500mm pipe diameter with UP04AST, please select
Z-path method.
Note3:
Do not use W-path method with UP04AST.
Note:
Do not select
“User Defined” item.
Summary of Contents for UFP-20
Page 1: ...Doc No KF08 002L Portable Ultrasonic Flowmeter UFP 20 Installation Operation Manual ...
Page 3: ... 2 The attachment of warning label is as follows Fig 1 2 Rear view of main unit Warning label ...
Page 8: ...1 Installation ...
Page 9: ......
Page 17: ......
Page 18: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 130: ...2 Operation ...
Page 131: ......
Page 182: ...3 Other ...
Page 183: ......
Page 193: ...3 8 3 2 4 Dimensions 1 Main Unit Fig 3 2 4 1 Main Unit Hand Strap ...