1-94
When main unit calculates transducer distance as -14 mm. Slide distance will be 36 mm.
So if up side slide sets at 100 mm point, DN side slide most be set at 136 mm point. The point of
scale is just sample. Whenever slide distance can be kept, scale point does not matter.
Fig. 1.2.9-23 Installation Sample of Z-path method (indicated value on main unit is -14 mm)
Slide distance
50
-14
36
Down side slide
Medium transducer
mounting fixture
Flow
Up side slide
Adaptor B
Reference point of DN
transducer at 136 mm point
Reference point of UP
transducer at 100 mm point
slide of Adaptor A
Reference point at
distance
Transducer
Flow
Adaptor A