3-5
Display
Method
LCD
(
320 x 240
Dot Matrix
)
/ high-intensity Backlight equipped
Content
• Instantaneous flow rate, warnings and totalizing status
• Forward/Backward flow totalized value
Digits
Flow rate
Max. 6 digits (including Sign section)
Flow velocity
Max. 6 digits
Sign section ; 1 digit
Integer section ; 2 digits
Decimal fraction ; 3 digits
Flow Totalizing
Max. 8 digits
Temp.
Max. 5 digits
Sign section ; 1 digit
Integer section ; 3 digits
Decimal fraction ; 1 digit
Unit
Flow rate units
m
3
/s,m
3
/min,m
3
/h,m
3
/D,km
3
/s,km
3
/min,km
3
/h,km
3
/D,Mm
3
/D
L/s,L/min,L/h,L/D
ft
3
/s,ft
3
/min,ft
3
/h,ft
3
/D,Mft
3
/D,bbl/s,bbl/min,bbl/h,bbl/D,Mbbl/D
gal/s,gal/min,gal/h,gal/D,Mgal/D,acf/s,acf/min,acf/h,acf/D,Macf/D
kg/s,kg/min,kg/h,kg/D,t/s,t/min,t/h,t/D,kt/s,kt/min,kt/h,kt/D,Mt/D
W,kW,MW,BTU/h,kBTU/h,MBTU/h
Totalize units
x1 m
3
,x5 m
3
,x10 m
3
,x100 m
3
x0.01 L,x0.1 L,x1 L,x10 L,X100 L
ft
3
,kft
3
,Mft
3
,bbl,kbbl,Mbbl,gal,kgal,Mgal,acf,kacf,Macf
x1 kg,x10 kg,x100 kg,x0.1 kg,x0.01 kg,x1 t,x10 t,x100 t
J,MJ,BTU,MBTU
Updating cycle
Approx. 1sec
Logging
Function
St’d/option
Standard
Output
Approx. 165,000 points
Date, time, Instantaneous flow rate (volumetric, mass or heat), + Total, -Total, Flow
velocity, Temperature, Error code (Selectable)
Internal logged data transferred through USB memory as CSV format
Output format
CSV
Temperature
Input
St’d/option
Option / Junction Box Required
Input
4pcs of Pt100 (Max.)
(For Energy measurement, they can be connected main unit through junction box.)
Analog Output
St’d/option
Standard
Output
1 port;
Instantaneous flow rate
Energy,
Mass (calculated by fixed density setting),
Calculation flow rate, mass or energy (path1 + path2 or path1 - path2)
Output format
4-20 mA
Allowable load resistance 550 ohm Max.
Summary of Contents for UFP-20
Page 1: ...Doc No KF08 002L Portable Ultrasonic Flowmeter UFP 20 Installation Operation Manual ...
Page 3: ... 2 The attachment of warning label is as follows Fig 1 2 Rear view of main unit Warning label ...
Page 8: ...1 Installation ...
Page 9: ......
Page 17: ......
Page 18: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 130: ...2 Operation ...
Page 131: ......
Page 182: ...3 Other ...
Page 183: ......
Page 193: ...3 8 3 2 4 Dimensions 1 Main Unit Fig 3 2 4 1 Main Unit Hand Strap ...