1-88
(3) Installation for medium transducer (Z-path method)
Sample installation Fig.1.2.9-17 is example as transducer distance as “-14mm”. The Z-path
method is single-traverse, the transducers are mounted diametrically opposite each other.
Please refer to chapter “
1.2.5 Selection of transducer setup position
(p.1-26)” for
recommendation place for transducer installation.
Fig. 1.2.9-17 Sample of Z-path method installation (Transducer distance : -14mm)
Note: Applicable pipe diameter for Z-path method installation by medium
transducer is DN20mm ~ 40mm.
Medium transducer
mounting fixture
Flow
Adaptor B
Slide distance
36
-14
Transducer
(Up side)
slide of Adaptor A
Reference point at
Adaptor A
50
Flow
Transducer
(Down side)
distance
Transducer