2-42
4. Alarm output
Following setting can be selected in case of alarming (R / D warning).
Table 2.2.8-9 Selectable items for alarm output
Alarm output
Remarks
Hold
Hold measured value
4 mA
Keep output 4mA
20 mA
Keep output 20mA
5. Analog output calibration
Here you can calibrate analog output on loop.
[Required item] Calibrated a
mpere meter
Fig. 2.2.8-8 Analog output calibration menu
a. Calibration on 4mA
When push
Select
key (F3 button), the main unit output 4mA
on a mandatory basis.
Input indicated mA value on the ampere meter.
b. Calibration on 20mA
When push
Select
key (F3 button), the main unit output 20mA
on a mandatory basis.
Input indicated mA value on the ampere meter.
A n a l o g O u t p u t C a l i b r a t i o n
1 :
4 m A O u t p u t
4 . 0 0 0 m A
2 : 2 0 m A O u t p u t
2 0 . 0 0 0 m A
# 0 1 : 1 0 0 / S t e e l / 1 M
B a c k
S e l e c t