Fig. 1.2.9-50 Set the mounting fixture with the magnets
4B-g. Add silicone grease onto surface of transducers and set into mounting fixture.
Add silicone grease enough to avoid any air layer between pipe and transducer.
And set transducers into holder, then tighten by screws.
1. Add couplant onto transducer.
2. Attach transducer with fixture by screw to be fit
onto pipe.
Fig. 1.2.9-51 Set the transducer
Fig. 1.2.9-52 The transducers installation completed
2
Transducer distance must be
measured between red-lines.
1
2
1. Attach fixture by magnet.
2. Check red line on the fixture of the side whether
positions properly.
1