![Tokyo Keiki UFP-20 Installation & Operation Manual Download Page 81](http://html1.mh-extra.com/html/tokyo-keiki/ufp-20/ufp-20_installation-and-operation-manual_1125190081.webp)
1-70
10. Thickness of lining
Input lining thickness by numeric button directly. Here for example, input “1.00mm”, then push
“
Enter
” key (F3 button) to proceed to next step.
Fig. 1.2.8-15 Lining thickness input menu
11. Fluid Selection
Select fluid from default choices or User Defined by direction or numeric button.
Here for example, select “
1:Water
”, then push “
Select
” key (F3 button) to proceed next step.
Fig. 1.2.8-16 Fluid type selection
After you select material, you will see predefined sound speed and viscosity, normally just
proceed to next. If you would like to select any un-listed fluid, please select “
User Defined
” then
enter actual sound speed of the fluid later at the next extra menu.
L i n i n g C o n d i t i o n
T h i c k n e s s
0 . 0
0
m m
# 0 1 :
1 0 0 / S t e e l / 1 M
B a c k
C l e a r
E n t e r
Push
Enter
key
F l u i d C o n d i t i o n
T y p e
1 :
W a t e r
2 : S e a w a t e r
3 : E t h y l e n e G l y c o l ( 5 0 w t % )
4 : G l y c e r i n
5 : A c e t o n e
6 : U s e r D e f i n e d
# 0 1 :
1 0 0 / S t e e l / 1 M
B a c k
S e l e c t
Push
Select
key
Note:
Over ½ of pipe diameter is invalid value. (Max range: up to 100mm)
Summary of Contents for UFP-20
Page 1: ...Doc No KF08 002L Portable Ultrasonic Flowmeter UFP 20 Installation Operation Manual ...
Page 3: ... 2 The attachment of warning label is as follows Fig 1 2 Rear view of main unit Warning label ...
Page 8: ...1 Installation ...
Page 9: ......
Page 17: ......
Page 18: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 130: ...2 Operation ...
Page 131: ......
Page 182: ...3 Other ...
Page 183: ......
Page 193: ...3 8 3 2 4 Dimensions 1 Main Unit Fig 3 2 4 1 Main Unit Hand Strap ...