![Teledyne LeCroy T3DSO2000A Series User Manual Download Page 99](http://html.mh-extra.com/html/teledyne/lecroy-t3dso2000a-series/lecroy-t3dso2000a-series_user-manual_1080819099.webp)
T3DSO2000A Manual. Page 95
In the list view:
•
Time— The horizontal offset of the current data frame head relative to the trigger
position.
•
Type— Type of the frame. Standard CAN frame is represented by "Std", CAN FD
frame is represented by "FD", the extended frame is represented by "Ext" and the
remote frame is represented by "RTR".
•
ID— Frame ID.
•
Length— Data length.
•
Data— Data bytes.
•
CRC— Cycle redundancy check.
•
Ack— Acknowledge bit.
16.9 I2S Trigger and Serial Decode
This section covers triggering and decoding I2S signals. Please read the following for
more details: " I2S Signal Settings", " I2S Trigger" and " I2S Serial Decode".
16.9.1 I2S Signal Settings
Connect the WS, BCLK, and Data signals to the oscilloscope set the mapping
relation between channels and signals, and then set the threshold level of each
signal. The process of specifying the source and threshold is similar to " I2C Signal
Settings"
BCLK
In addition to specifying the source and the threshold level, BCLK signals also require
the specification of the Edge Select.
•
Rising– Data latched on the rising edge of the clock.
•
Falling– Data latched on the falling edge of the clock.
www.