HFR: Signals are DC coupled to the trigger circuit and a low-pass filter network
attenuates high frequencies (used for triggering on low frequencies). See the
datasheet for details of the cut-off frequency.
•
LFR: The signal is coupled through a capacitive high-pass filter network, DC is rejected
and low frequencies are attenuated. For stable triggering on medium to high-frequency
signals. See the datasheet for details of the cut-off frequency.
15.9 Noise Reject
Noise Reject adds additional hysteresis to the trigger circuitry. By increasing the trigger
hysteresis, the noise immunity becomes better but the trigger sensitivity degrades.
Noise Reject Off
www.
.com
1.800.561.8187