T3DSO2000A Manual. Page 166
A. Set the scale of the amplitude coordinate axis
B. Set the reference level, the max value of the amplitude
coordinate axis
C. Set the amplitude mode. Set to Vout to display the
amplitude value of the output signal; set to Vout/Vin to
display the amplitude ratio of the output signal to the input
signal
D. When the mode is Vout, it needs to set the unit (Vpp, Vrms,
dBV, dBu, dBm or Arbitrary dB); when the mode is
Vout/Vin, it needs to set the amplitude axis type (Linear or
Logarithmic).
E. Auto Set. According to the output signal amplitude curve,
the oscilloscope automatically sets the scale and reference
level
F. Return to the previous menu
Phase
Set the phase coordinate axis of the bode plot. Touch
Display>Phase to recall the phase setting dialog box:
A. Set the scale of the phase coordinate axis
B. Set the reference level, the max value of the phase
coordinate axis
C. Set the phase unit (Degree or Rad)
D. Auto-Set. According to the output signal phase curve, the
oscilloscope automatically sets the scale and reference level
E. Return to previous menu
Cursors
The T3DSO2000A can use the cursor to measure the bode plot curve. The cursor of
the bode plot is similar to the ordinary cursor, see the chapter "Cursors" for details.
Touch Display>Cursors to recall the cursor setting dialog box.