T3DSO2000A Manual. Page 79
If you set the trigger condition to 7 address & data or 10 address & data:
•
Address can be selected in the hexadecimal range of 0x00 to 0x7F
(7-bit) or 0x3FF (10-bit). If the address is selected as "0xXX (7-bit
address)" or "0xXXX (10-bit address)", the address is ignored. It
will always trigger on the Ack bit followed by the address.
•
Data1 and Data2 can be selected in the hexadecimal range of
0x00 to 0xFF. If the data is selected as "0xXX", the data is ignored.
It will always trigger on the Ack bit followed by the address.
•
R/W Bit can be specified as Write, Read or Don't Care.
Data Length
— The data length range is 1 to 12 bytes. When the
SDA data length is equal to the value of the Data Length setting and
the address length is the same as the set value, the oscilloscope will
be triggered.
•
Touch the Address Length to select "7-bits" or "10-bits" to match the address of the
input signal.
•
Touch the Data Length and turn the universal knob or by virtual keypad to set the
data length to match the data length of the input signal.