T3DSO2000A Manual. Page 52
A. Touch the level region and rotate the Level knob on the front panel to adjust trigger
level; touch the region again and the virtual keypad will be available
B. ▲ to increase and ▼ to decrease the trigger level
C. Set trigger level to the vertical center of the waveform
automatically
D. Set the trigger mode to "Auto", which is equivalent to
pressing the Auto button on the front panel
E. Set the trigger mode to "Single", which is equivalent to
pressing the Single button on the front panel
F. Set the trigger mode to "Normal", which is equivalent to
pressing the Normal button on the front panel
A. Select trigger type
B. Select the trigger source
C. Select the trigger slope (When the trigger type is "Edge",
"Slope" and other specific types)
D. Set holdoff (None/Time/Events)
E. Set trigger coupling mode (DC/AC/LF Reject/HF Reject)
F. Enable/disable Noise Rejection. When Noise Reject is on,
the trigger hysteresis is increased, so the noise immunity
of the trigger circuit is better. As a compromise, the trigger
sensitivity degrades
G. Set the Zone trigger