T3DSO2000A Manual. Page 115
•
FRLR
:
The time between the first rising edge of source A and the last rising edge of
source B at the 50% crossing
•
FRLF
:
The time between the first rising edge of source A and the last falling edge of
source B at the 50% crossing
•
FFLR
:
The time between the first falling edge of source A and the last rising edge of
source B at the 50% crossing
•
FFLF
:
The time between the first falling edge of source A and the last falling edge of
source B at the 50% crossing
•
Skew: Time of source A edge minus time of nearest source B edge
18.4 Trend
After adding a measurement parameter, Trend can be used to observe the change of
the selected measurement value over time.
A. Measurement parameter display area
B. Trend plot display area
C. Extend the range of time. Touch it to expand the time range.
D. Current value
E. Maximum value
G. Average value
F. Minimum value
H. Reset statistics
Press Clear Sweeps button or touch Reset Statistics in the measure dialog box, or touch
the symbol in the statistics display area to clear and restart statistics.