T3DSO2000A Manual. Page 34
The oscilloscope will automatically convert the vertical scale according to the current
probe attenuation factor. For example, the vertical scale of the oscilloscope under 1X
attenuation is 100 mV/div, and the vertical scale will be automatically set to 1 V/div if the
probe attenuation is changed to 10X. If a standard probe with readout terminal is
connected, the oscilloscope will automatically set the probe attenuation to match the
probe.
Label
Users can set the label text for channels. Open the label setting dialog box to select the
source, customize the label text, and set the display. The source can be C1~C4, F1~F2,
and RefA~RefD. The length of the label is limited to 20 characters. The characters
beyond this length will not be displayed. When the “Display” option is set to “on”, the
label will be displayed on the right side of the channel offset indicator.
The labels are hidden
The displayed labels
Applied to
With this setting, some common functions such as Trigger, Cursor, Measure, FFT,
Search, Mask Test, Counter, and AWG can be quickly applied to the selected channel.
Once a function is specified, it will switch directly to the function menu and automatically
set that channel as the source.
Impedance
•
1 MΩ: When a passive probe with high impedance is connected, the impedance must
be set to 1 MΩ, otherwise the signal will not be detected.
•
50 Ω: Suitable for high-frequency signals transmitted through 50 Ω coaxial cables and
can minimize the amplitude distortion caused by impedance mismatching.
Unit
Set the vertical scale units to "V" or to "A”. When using a voltage probe, the units should
be set to "V". When using a current probe, the units should be set to "A".