![Teledyne LeCroy T3DSO2000A Series User Manual Download Page 39](http://html.mh-extra.com/html/teledyne/lecroy-t3dso2000a-series/lecroy-t3dso2000a-series_user-manual_1080819039.webp)
T3DSO2000A Manual. Page 35
Deskew
Due to the skew between channels, cables or probes, the delay of signals passing
through different measurement paths may be inconsistent. For example, two coaxial
cables with a 1-inch difference in length could introduce a skew of more than 100 ps. In
some scenarios (e.g. measuring the setup/hold time between clock and data), it may be
necessary to compensate the skew between channels.
The method of compensation: Probe the same signal simultaneously using two channels
(including the cables or probes that you intend to use for measurements) and adjust the
deskew parameter of one channel until the waveforms of the two channels observed on
the screen coincide horizontally.
Before deskew
After deskew
Note
: When the channel with deskew compensation is the trigger source, the trigger
delay indicator does not change as the deskew value changes.
Invert
When invert is enabled, the waveform is displayed 180 degrees opposite to the earth
potential. This is a mathematical inversion and does not physically change the actual
potential of the input signal.
Before Invert
After Invert
www.