16.2 I2C Trigger and Serial Decode
This section covers triggering and decoding I2C signals. Please read the following for
more details: " I2C Signal Settings", " I2C Trigger" and " I2C Serial Decode".
16.2.1 I2C Signal Settings
Connect the serial data signal (SDA) and the serial clock signal (SCL) to the oscilloscope,
set the mapping relation between channels and signals and then set the threshold level of
each signal. The signal settings of decoding and triggering are independent. If you want to
synchronize the settings between decode and trigger, please perform Copy Setting in the
decode dialog box.
A. Set the source of SCL. In the example above, SCL is connected to C4.
B. Set the threshold level of SCL. It is 1.7V for the LVTTL signal in this example.
C. Set the source of SDA. In the example above, SDA is connected to C1.
D. Set the threshold level of the SDA channel.
E. Return to the previous menu.
F. Threshold level line. It only appears when adjusting the threshold level.
Copy Setting Touch the Copy Setting in the decode dialog box to synchronize the settings
between the trigger and decoding configurations.
A. Copy the decode settings to trigger
B. Copy the trigger settings to decode
C. Return to the previous menu.