User Manual
PCS 200 Kit
- 3.37 -
3.8
ALARMS
The alarms and warnings are reported in a pop-up window like the one shown in the
picture.
The
➌
button confirms the alarm acknowledgment and closes the window, but it
does not cancel the alarm.
There are various kinds of alarms. According to the kind of alarm, it is possible to
cancel it (the alarm status is eliminated) automatically or through a system rebooting,
pressing the battery disconnect switch.
Category
Background
colour
Buzzer
The pop-up
window closes
Description
Unrecoverable
emergency
Red
5 continuously
repeated bips
Manual,
pressing
the
alarm
acknowledgment
button
It is not possible to go on
seeding.
The system must be
disconnected from the
power supply
Recoverable
emergency
It is not possible to go on
seeding.
The system must be
disconnected from the
power supply
Standard error
1 bip
It is possible to go on
seeding but the device
showing an error is not
working properly
Warning
Yellow
1 bip
Automatic after
3 seconds
It is possible to go on
seeding but the operator
should be aware of the
situation.
The already checked (acknowledgment button pressed) but still active errors, will be
visualized in the window in a lighter background colour.
Summary of Contents for PCS 200
Page 1: ...PCS 200 Installation Operation Maintenance ORIGINAL INSTRUCTIONS ...
Page 4: ......
Page 33: ...Installation PCS 200 Kit 2 13 Figure 2 Wiring diagram of the system without alternator ...
Page 35: ...Installation PCS 200 Kit 2 15 Figure 2 14 ECU wiring diagram ...
Page 41: ...Installation PCS 200 Kit 2 21 This page is intentionally left blank ...
Page 102: ...User Manual PCS 200 Kit 3 61 This page is intentionally left blank ...
Page 105: ...Maintenance and Troubleshooting PCS 200 Kit 4 3 4 4 SPARE PARTS ...
Page 111: ...Maintenance and Troubleshooting PCS 200 Kit 4 9 This page is intentionally left blank ...