Firmware Update and Installation of Firmware Options
Firmware Update
3.2
Quick Start Guide 1313.9652.62 - 02
3.1 Firmware Update
For the latest firmware version refer to the Rohde & Schwarz internet site and download the
most up-to-date firmware.
A detailed description about performing the firmware update in the intrument is given in the
release note, provided on this CD. The release note is also downloadable from the
Rohde & Schwarz internet site.
You can install a new firmware version by using a memory stick, a USB CD-ROM, or LAN. The
firmware update kit contains several disks whose contents must be stored in separate
directories (Disk1 to Disk<n>) when a memory stick or LAN is used. Start the installation
program from the
SETUP
menu.
➢
Press the
SETUP
key.
The
SETUP
menu will open.
➢
Press the
NEXT
key.
The side menu will open.
➢
Press the
FIRMWARE UPDATE
softkey.
The subdirectory for installing or uninstalling new firmware versions will open.
➢
Press the
UPDATE PATH
softkey. The entry field for the path with the set of update
directories will open.
➢
Enter the drive name and directory and confirm with
ENTER
.
Example:
The firmware update files are stored on a memory stick in the
Update
directory. They are
allocated in the
Disk1
to
Disk<n>
subdirectories. After you insert the memory stick, it will be
detected as drive
E:
.
In this case, the required path specification is
E:\UPDATE
.
➢
To start the installation program, press the
FIRMWARE UPDATE
softkey in the submenu.
➢
Confirm the query
DO YOU REALLY WANT TO UPDATE THE FIRMWARE?
with
OK
.
The installation program will then guide you through the remainder of the update procedure.
The instrument will perform several automatic shutdowns, until the new base system
firmware is installed properly.
IEC bus command:
"SYST:FIRM:UPD 'E:\UPDATE'"
➢
To restore the previous firmware version, press the
RESTORE FIRMWARE
softkey.
IEC bus command:
--
Do not switch the instrument off until the update process has been finished completely.
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...