R&S FSQ
Measurements in the Time Domain
5.22
Quick Start Guide 1313.9652.62 - 02
4. Set the sweep time to 1 ms.
➢
Press the
SWEEP
key and enter
1 ms
.
The R&S FSQ will show the GSM burst continuously across the display.
5. By using the video trigger, elicit triggering on the rising edge of the burst.
➢
Press the
TRIG
key.
➢
Press the
VIDEO
softkey and enter
70%
.
The R&S FSQ shows a static image with the GSM burst at the start of the trace. The
trigger level is displayed as a horizontal line labelled with the absolute level for the trigger
threshold in the measurement diagram.
6. Configure power measurement in the time domain.
➢
Press the
MEAS
key.
➢
Press the
TIME DOM POWER
softkey.
➢
Set the
LIMITS
softkey to
ON
.
➢
Press the
START LIMIT
softkey.
➢
By turning the rotary knob clockwise, move the vertical line to the start of the burst.
➢
Press the
STOP LIMIT
softkey.
➢
By turning the rotary knob counterclockwise, set the second vertical line to the end of the
burst.
The R&S FSQ will output on screen the average (mean) power during the activation
phase of the burst.
Fig. 5-11 Measurement of the average power during the activation phase of a GSM
burst.
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...