R&S FSQ
Storing and Loading Instrument Settings
5.36
Quick Start Guide 1313.9652.62 - 02
5.6.2.1 Automatic Loading of a Data Record during Booting
If the R&S FSQ is switched on in the factory default state, it loads the instrument settings that
it had when switched off (provided that it was switched off using the
STANDBY
switch on the
“ Switching Off the R&S FSQ” on page 2.8
).
However, the R&S FSQ can also automatically load a data record defined by the user. This
requires performing the following procedure:
➢
Press the
FILE
key and then the
RECALL
softkey.
➢
Press the
STARTUP RECALL
softkey.
The list of available data records is shown.
➢
Select the data record to be loaded by using the rotary knob and mark it with
ENTER
.
➢
Close the dialog box by pressing
ESC
twice.
• The selected data record will also be loaded by pressing the
PRESET
key.
• The
FACTORY
entry activates factory-default operation, i.e. the
settings that were present at the time of deactivation will be loaded
at activation.
• If you need to change the path for the instrument configuration, do
so by using the
EDIT PATH
softkey.
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...