![Rohde & Schwarz FSQ series Quick Start Manual Download Page 195](http://html.mh-extra.com/html/rohde-and-schwarz/fsq-series/fsq-series_quick-start-manual_1477838195.webp)
Quick Start Guide 1313.9652.62 - 02
7.13
R&S FSQ
Brief Introduction to Remote Control
Detailed Programming Examples
7.2.1.2 Default Settings for Measurements
REM ************************************************************************
Public Sub SetupInstrument()
Dim status As ViStatus
Dim retCnt As Long
REM ------ Default setting of the R&S FSQ ----------------------------------
Call SetupStatusReg
'Configure status registers
status = viWrite(analyzer, "*RST", 4, retCnt)
'Reset instrument
status = viWrite(analyzer, "SYST:DISP:UPD ON", 16, retCnt)
'ON: screen display on, OFF: off(improved performance)
status = viWrite(analyzer, "DISP:FORM SINGle", 16, retCnt)
'Full screen display
status = viWrite(analyzer, "DISP:WIND1:SEL", 14, retCnt)
'Active Screen A
status = viWrite(analyzer, "INIT:CONT OFF", 13, retCnt)
'Single sweep mode
REM ------ Frequency setting ----------------------------------------------
status = viWrite(analyzer, "FREQUENCY:CENTER 100MHz", 23, retCnt)
'Center frequency
status = viWrite(analyzer, "FREQ:SPAN 1 MHz", 15, retCnt)
'Span
REM ------ Level setting --------------------------------------------------
status = viWrite(analyzer, "DISP:WIND:TRAC:Y:RLEV -20dBm", 28, retCnt)
'Reference level
status = viWrite(analyzer, "INP:ATT 10dB", 12, retCnt)
'Input attenuation (*)
REM ------ Level scaling --------------------------------------------------
status = viWrite(analyzer, "DISP:WIND:TRAC:Y:SPAC LOG", 25, retCnt)
'Log level axis
status = viWrite(analyzer, "DISP:WIND:TRAC:Y:SCAL 100dB", 27, retCnt)
'Level range
status = viWrite(analyzer, "DISP:WIND:TRAC:Y:SCAL:MODE ABS", 30, retCnt)
'Absolute scaling
status = viWrite(analyzer, "CALC:UNIT:POW DBM", 17, retCnt)
'y meas. unit
REM ------ Trace and detector setting -------------------------------------
status = viWrite(analyzer, "DISP:WIND:TRAC1:MODE AVER", 25, retCnt)
'Trace1 average
status = viWrite(analyzer, "AVER:TYPE VID", 13, retCnt)
'Average mode video; "VID" for video
status = viWrite(analyzer, "SWE:COUN 10", 11, retCnt)
'Sweep count
status = viWrite(analyzer, "DISP:WIND:TRAC2:STAT OFF", 24, retCnt)
'Trace2 blank
status = viWrite(analyzer, "DISP:WIND:TRAC3:STAT OFF", 24, retCnt)
'Trace3 blank
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...