Quick Start Guide 1313.9652.62 - 02
2.5
R&S FSQ
Preparing for Use
Preparing for Operation
2.1.2 Setting Up the Instrument
The instrument is intended for indoor use. Note the following in this regard:
• The ambient temperature must comply with the range specified in the data sheet.
• Air inflow and outflow via the air vents on the rear panel and through the lateral perforations
must not be obstructed. Clearance from walls must be at least 10 cm.
• The mounting surface must be even.
Risk of damaging electronic components
To avoid damage of electronic components, the operational site must be protected against
electrostatic discharge (ESD).
The following two methods of ESD protection may be used together or separately:
• Wrist strap with cord to ground connection
• Conductive floor mat and heel strap combination
Wrist strap with cord
Building ground
Ground connection
of operational site
Floor mat
Heel strap
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...