![Rohde & Schwarz FSQ series Quick Start Manual Download Page 113](http://html.mh-extra.com/html/rohde-and-schwarz/fsq-series/fsq-series_quick-start-manual_1477838113.webp)
Quick Start Guide 1313.9652.62 - 02
5.3
R&S FSQ
Measuring a Sinusoidal Signal
5.1 Measuring a Sinusoidal Signal
One of the most common measurement tasks is determining the level and frequency of a
signal. When measuring an unknown signal, you can usually start with the preset setting.
In the following examples, a signal generator is used as a signal source.
Test setup:
Connect the RF output of the signal generator to the RF input of R&S FSQ.
Settings on the signal generator:
5.1.1 Measuring the Level and Frequency with Markers
The level and frequency of a sinusoidal signal can be measured easily by using the marker
function. The R&S FSQ always displays its amplitude and frequency at the marker position.
The frequency measurement uncertainty is determined by the reference frequency of the
R&S FSQ, the resolution of the marker frequency display and the resolution of the screen.
5.1.1.1 Procedure
1. Reset the instrument.
➢
Press the
PRESET
key.
2. Connect the signal to be measured to the RF INPUT input on the instrument front
panel.
3. Set the center frequency to 128 MHz.
➢
Press the
FREQ
key.
The entry field for the center frequency appears on screen.
➢
Using the numeric keypad in the input field, enter
128
and confirm the entry with the
MHz
key.
If levels higher than +30 dBm (=1 W) are expected or are possible, a
power attenuator must be inserted before the RF input of the R&S FSQ.
If this is not done, signal levels exceeding 30 dBm can damage the RF
attenuator or the input mixer. The total power of all signals that are
present must be taken into account.
Frequency:
128 MHz
Level:
-30 dBm
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...