![Rohde & Schwarz FSQ series Quick Start Manual Download Page 45](http://html.mh-extra.com/html/rohde-and-schwarz/fsq-series/fsq-series_quick-start-manual_1477838045.webp)
Quick Start Guide 1313.9652.62 - 02
1.17
R&S FSQ
Front and Rear Panel
Rear Panel Connections
RBW 20 MHz and 50 MHz
Output voltage at signal level / dBm - RF_Attenuation / dB = 0:1.8 V
Slope:14 mV / dB
Exception: Ref_Level / dBm - RF_Attenuation / dB > 0
Output voltage at signal level / dBm - RF_Attenuation / dB = 0
1.8 V - (Ref_Level / dBm - RF_Attenuation / dB) * 14 mV
TG I / AM IN; TG Q / FM IN (Option R&S FSU-B9)
The two female connectors TG I /AM IN and TG Q /FM IN are used to modulate the tracking
generator (option R&S FSU-B9) by means of an external signal.
The input voltage range is ±0.5 V; the input impedance is 50
Ω
.
LAN Interface
The LAN interface can be used to connect to a local network. The assignment of the RJ-45
connector supports twisted-pair category 5 UTP/STP cables in a star configuration (UTP
stands for “unshielded twisted pair”, and STP for “shielded twisted pair”).
Second GPIB Interface IEC2 (Option R&S FSP-B10)
When equipped with option R&S FSP-B10 (external generator control), the instrument
provides a second GPIB interface for using external generators.
Fig. 1-8
Pin assignment of the second IEC interface
The R&S FSQ remote control is not supported via the second GPIB
interface.
12
1
13
24
SHIELD SRQ NDAC DAV DIO4 DIO2
LOGIC GND GND(10) GND(8) GND(6) DIO8 DIO6
GND(11) GND(9) GND(7) REN DIO7 DIO5
ATN IFC NRFD EOI DIO3 DIO1
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...