333
16
Delivery Rate Accuracy
16.6.2 Trumpet Curve
The trumpet curve shows the delivery rate accuracy during the
observation period.
0,1 U/h, 6 mm cannula
-92,0000
-43,0000
-14,4000
-3,6364
-0,3158
1,4839
204,0000
64,0000
24,8000
14,3636
9,7895
7,6129
4,5200
-150
-100
-50
0
50
100
150
200
250
0
60
120
180
240
300
360
420
480
540
600
660
720
780
840
900
960
Observation period [min]
Error as percentage [%]
Minimum flow error
Maximum flow error
Total error
man_07866364001_C_en-UK.indb 333