Maintenance Messages
15
Code
Title of maintenance
message
Possible cause/consequences
Further information
M-59
Battery almost empty
The level of the rechargeable battery is
very low.
The diabetes manager automatically deactivates
communication via
Bluetooth
wireless technology
to save power. As a result, communication with the
micropump is interrupted. Recharge the battery of
your diabetes manager.
M-60
Clock error
A discrepancy in the internal clocks of
the micropump system was detected.
Set the current time and the current date on the
diabetes manager.
M-62
Connection failed
The pairing code was not scanned
successfully. This may be the case, for
example, if it is too dark or if the code
or camera lens is dirty and the code
cannot be read correctly.
Try rescanning the pairing code on the micropump.
Alternatively, you can enter the pump key manually.
M-64
Bolus delivery not
possible
The connection between the diabetes
manager and micropump was lost.
Hold the diabetes manager close to the micropump
and ensure that data communication is not
disrupted.
You can deliver a quick bolus straight from the
micropump.
M-65
Bolus delivery not
possible
The micropump is in STOP mode.
If you want to deliver a bolus, start the micropump
first.
man_07866364001_C_en-UK.indb 307
12.10.2018 13:10:33
Summary of Contents for Accu-Chek Solo
Page 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Page 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Page 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Page 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Page 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...