4
69
Putting the Components into Operation
1
4
3
Disinfect the top of the insulin
vial with a sterile alcohol wipe.
Allow the disinfected top of the
insulin vial to dry.
Hold the round part of the
handle firmly and pull it
downwards in the direction of
the arrow in order to fill the
reservoir with air.
Fill the reservoir with the
volume of air that you later
want to fill with insulin.
Carefully pull out the battery’s
protective film downwards in
the direction of the arrow to
activate the battery.
2
Remove the new reservoir
assembly from the packaging.
man_07866364001_C_en-UK.indb 69
12.10.2018 13:08:10