Error Messages
15
Code
Title of error
Possible cause/consequences
Possible solutions
Error messages triggered by the micropump.
E-6
Mechanical error in the
micropump
The micropump switches to STOP
mode and does not deliver any insulin.
Select the
Replace
menu and replace the
reservoir. If the problem persists, replace the
micropump.
E-7
Electronic error
Communication between the
micropump and diabetes manager is
not possible. The micropump switches
to STOP mode and does not deliver
any insulin.
Select the
Replace
menu and replace the
reservoir. Wait at least 30 seconds after removing
the used reservoir before connecting a new
reservoir to the pump base. If the problem
persists, replace the micropump.
E-8
Micropump battery error
The energy supply is defective.
The micropump switches to STOP
mode and does not deliver any insulin.
After 10 seconds the pump turns off.
Select the
Replace
menu and replace the
reservoir.
man_07866364001_C_en-UK.indb 311
12.10.2018 13:10:34
Summary of Contents for Accu-Chek Solo
Page 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Page 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Page 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Page 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Page 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...