Putting the Micropump into Operation
w
WARNING
j
Check all components for visible damage before using them
with the micropump system.
j
The reservoir, cannula and pump holder are intended for
single use and are sterile packaged. They must not be used
if their sterile packaging was previously opened or damaged
or if the use by date has expired.
j
If the infusion assembly may have come into contact with
infectious material, replace the infusion assembly
immediately. There is a risk that infections (for example,
hepatitis or HIV) could be transmitted.
j
If there is an unexpected rise in your blood glucose level or
an occlusion message occurs, check the micropump and
the infusion assembly for occlusions and leaks. Replace
your infusion assembly if you are not sure whether the
infusion assembly is working properly.
j
Check regularly to make sure that the pump holder does not
detach itself from the infusion site and that the adhesive pad
is not wet. Insulin delivery may be interrupted by a loose
fitting or displaced cannula.