Maintenance Messages
15
Code
Title of maintenance
message
Possible cause/consequences
Further information
M-87
Micropump not stopped The micropump cannot be stopped.
Try to stop the micropump again. If the pump does
not stop, remove the micropump from your body,
switch to an alternative therapy method and
contact your Customer Support and Service Centre.
M-88
Flight mode turned off
The micropump and the diabetes
manager are not within communication
range. Flight mode could not be turned
on on the micropump.
Hold the diabetes manager close to the micropump.
M-94
Connection failed
There is a communication problem
between the micropump and the
diabetes manager.
Hold the diabetes manager close to the micropump
and ensure that data communication is not
disrupted.
Check the micropump.
M-95
No micropump located
A connection to the micropump could
not be established.
Check whether the micropump is too far away and
restart the locating process.
M-96
USB connection failed
The USB connection between the
diabetes manager and the software on
the computer failed.
Check whether the software is correctly installed
on the computer.
man_07866364001_C_en-UK.indb 309
12.10.2018 13:10:33
Summary of Contents for Accu-Chek Solo
Page 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Page 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Page 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Page 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Page 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...