Messages and Troubleshooting
Code
Title of maintenance
message
Possible cause/consequences
Further information
Maintenance messages triggered by the diabetes manager.
M-51
Test strip error
The test strip is used, damaged or not
completely inserted into the test strip
slot.
Use a new test strip or re-insert the test strip into
the test strip slot.
M-53
Test failed
The blood glucose test did not work
properly.
Repeat the blood glucose test with a new test strip.
M-54
Drop too small
The amount of blood or control solution
is not sufficient to perform a test.
Repeat the test with a new test strip. Make sure the
blood drop or drop of control solution is large
enough.
M-56
Drop applied too early
The drop was drawn into the test strip
before
Apply drop
appeared on the
screen.
Repeat the test with a new test strip and a fresh
blood drop or drop of control solution.
M-58
Temperature too high or
too low
The ambient temperature for testing
blood glucose or performing a control
test is outside the permitted range.
Make sure the ambient temperature is within the
permitted range.
Wait 5 minutes before testing your blood glucose
again or performing a control test.
man_07866364001_C_en-UK.indb 306
12.10.2018 13:10:33
Summary of Contents for Accu-Chek Solo
Page 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Page 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Page 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Page 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Page 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...