Warnings
15
Code
Title of warning
Information/possible cause
Information/possible solution
W-86
Flight mode on
Data cannot be synchronised between the
diabetes manager and the micropump
because flight mode is turned on.
Therefore, the entries saved for the bolus
advice may not be up to date. The diabetes
manager can currently only give bolus
advice based on the entries saved in the
diabetes manager.
You can still use the bolus advice feature. Note,
however, that the diabetes manager is not
receiving any information (for example, errors)
from the micropump.
When flight mode is turned off and the diabetes
manager and micropump are within
communication range, the data will be
synchronised. The data for bolus advice will
then be up to date again.
W-88
Flight mode on
The saved logbook entries may not be up to
date.
When flight mode is turned off and the diabetes
manager and micropump are within
communication range, the logbook entries will
be synchronised.
W-89
Check logbook entries
It was not possible to assign a bolus
delivered by the micropump to a bolus
confirmed in the bolus advice function.
Correct the logbook entries as required.
W-90
Time synchronised with
micropump
The time difference between the diabetes
manager and the micropump was
corrected.
Check the time on the diabetes manager.
man_07866364001_C_en-UK.indb 301
12.10.2018 13:10:32
Summary of Contents for Accu-Chek Solo
Page 2: ...man_07866364001_C_en UK indb 2 12 10 2018 13 07 31 ...
Page 5: ...man_07866364001_C_en UK indb 5 12 10 2018 13 07 32 ...
Page 44: ...32 Getting to Know the Micropump System man_07866364001_C_en UK indb 32 12 10 2018 13 07 43 ...
Page 276: ...264 Reminders man_07866364001_C_en UK indb 264 12 10 2018 13 10 13 ...
Page 330: ...318 Messages and Troubleshooting man_07866364001_C_en UK indb 318 12 10 2018 13 10 35 ...