List Operation Function (List)
In list operation mode, you can perform the complex
current/voltage/resistance/power modes accurately and rapidly, which may be
synchronized with internal or external signals. In this way, you can complete the
sophisticated test.
In this mode, when receiving a trigger signal, the load will operate according to the
set execution mode and sequence. As shown in Figure 2-26, the load simulates the
complex sequences of input changes based on the list parameters that you edit. The
list function supports CC, CV, CR, and CP modes.
Figure 2-26 List Operation Function (List)
When selecting different trigger sources, you can generate complex sequences by
editing the setting value for each step, the dwell time, and slew rate (the slew rate
can only be edited in CC mode) to meet the complex test demands. All listed data is
stored in a non-volatile memory. It means that the data will not be lost when the load
is turned off.
Operation Procedures:
1.
Power on the instrument
Press the Power key
on the front panel to power on the instrument.
2.
Connect the DUT and the channel input terminals of the load
Connect the DUT and the channel input terminals on the front panel of the load
as shown in Figure 2-2.
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. A misconnection with the terminals may cause damage to
the instrument and the devices connected to it.
0
TRIG
Circle=1
Circle=2
0
TRIG
Circle=1
Circle=2
1 2 3 4 5
0
TRIG
Circle=1
Circle=2