Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-27
Set B_Level
The sink current toggles between a high value and a low value in Pul mode.
B_Level indicates a low value. The default unit for B_Level is Ampere (A).
(1)
In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "B_Level".
(2)
Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Parameter Setting Method".
Note: The set A_Level and B_Level for a specified model should be within its
specified high and low ranges for the model.
Set Slew_RIS
Sets the sink current slew rate for the positive transitions from Level B to Level A
in Pul mode. The default unit for the rising slew rate is A/μs.
(1)
In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "Slew_RIS".
(2)
Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Parameter Setting Method".
Set Slew_FAL
Sets the sink current slew rate for the negative transitions from Level A to Level
B in Pul mode. The default unit for the falling slew rate is A/μs.
(1)
In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "Slew_FAL".
(2)
Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions in "Parameter Setting Method".
Set Width
The time during which the sink current stays at Level A when it switches to Level
A in Pul mode. Its unit is s or ms.
(1)
In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "Width".
(2)
Enter a value by using the numeric keys. For the setting methods, refer to
the descriptions of the built-in help information or refer to the descriptions
in "Parameter Setting Method".
Set TRIG_SRC
Three triggers are available in Pul mode: BUS trigger, TRAN trigger, and DIGIO
trigger.
(1)
In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "TRIG_SRC".
(2)
Press the Left/Right arrow key to select the type of the trigger source.
4.
Enter the main interface
In the guide interface, after you configure the parameters, press Apply to enter
the main interface, as shown in Figure 2-20. At this time, the parameters are