Chapter 2 Front Panel Operations
RIGOL
DL3000 User’s Guide
2-67
(1)
In the main interface, press Range to switch the parameter focus to
"Range".
(2)
Press Range to switch the current working range. You can also use the
Left/Right arrow key or the knob to switch it.
If you select "CV" mode:
(1)
In the main interface, press Range to switch the parameter focus to
"Range".
(2)
Press Range to switch the voltage working range. You can also use the
Left/Right arrow key or the knob to switch it.
Note:
The low range provides better resolution and accuracy at low current or
voltage settings.
If the set current or voltage value is greater than the maximum value of the
low range, you must select a high range.
CAUTION
Before switching the current range, please disable the channel
input to avoid causing damage to the instrument or the DUT.
Set Max
Sets the upper limit of the working voltage or current in the current mode based
on the currently selected working mode for the factory test function.
(1)
In the main interface, press Max to switch the parameter focus to "Max".
(2)
Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to
the descriptions in "Parameter Setting Method".
Set Min
Sets the lower limit of the working voltage or current in the current mode based
on the currently selected working mode for the factory test function.
(1)
In the main interface, press Min to switch the parameter focus to "Min".
(2)
Enter a value by using the numeric keys, arrow keys, or the knob. For
setting methods, refer to the descriptions of the help information or refer to
the descriptions in "Parameter Setting Method".
Set Pass_Tip
If the test is passed, you can determine whether to display the prompt message
that reports the test pass.
(1)
In the main interface, press Pass_Tip to switch the parameter focus to
"Pass_Tip".
(2)
Press Pass_Tip to enable or disable the display of the prompt message.
You can also use the Left/Right arrow key or the knob to switch the on/off
state.